Abstract

This paper presents results of the processing and validation of data collected by an automatic gas chromato-graph (AGC). This system was used to monitor 62 volatile organic compounds (VOCs) in urban air in the Basque Country, Spain. The nonpolar compounds (C2–C10) identified—paraffins, olefins, aromatics, and chlorinated compounds—accounted for 88% of the mass of total non-methane hydrocarbons (TNMHCs) in ambient air. The evaluation of linearity, precision, detection limits (DLs), and stability of retention times (RTs) indicates that the equipment is suitable for measuring ambient air automatically for prolonged periods (6 months). The calibration of the equipment using response factors calculated on the basis of the effective carbon number (ECN) showed variations of over 10% for acetylene, isoprene, and n-hexane. The results provided by the automatic chromatograph correlated significantly with simultaneous results from other widely used techniques for determining VOCs in ambient air: (1) portable GC, equipped with photoionization detector (PID), and (2) active adsorption on Tenax-TA followed by thermal des-orption and chromatographic analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.