Abstract
ABSTRACTAn approach to the on-line coupling of solid-phase extraction on porous graphitic carbon (PGC) column and high performance liquid chromatography (HPLC) separation on octadecyl silica column is proposed. The approach includes extraction of phenols from 10 ml of water sample on PGC column; desorption with subcritical water at 175°C, injection of the eluate to the HPLC column and separation at ambient temperature with acetonitrile–water eluent. Limits of detection for UV detection of phenols were 1–5 μg l−1, and analyte peaks were 20–50% narrower than for direct injection of 20 μl of sample.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.