Abstract

The on-line coupling of gel permeation chromatography (GPC) and 1H NMR operating at temperatures up to 130 °C is presented. A NMR flow probe with a cell volume of 120 μL and a stop-flow valve are developed for on-flow and stop-flow NMR measurements at high temperatures. To maintain high and constant temperatures through the whole probe, the flow probe contains two separate heating circuits. A modified stop-flow valve is developed as a control device for enabling on-flow and stop-flow experiments at high temperature conditions. Heated transfer lines connect the flow probe with the high temperature GPC system. Due to their semicrystalline nature, polyolefins can be studied by liquid chromatography only at temperatures above 100 °C. The novel high temperature GPC–NMR system is used for the separation of complex polyolefins regarding their molar mass and for the analysis of different chemical structures. Blends of polyethylene, poly(methyl methacrylate), and ethylene-methyl methacrylate copolymers are separated according to the molar masses of the components. The compositions of the components are directly studied by on-line NMR. Moreover, the chemical composition distribution of an ethylene-methyl methacrylate copolymer sample is analysed. Differences between results of on-flow and stop-flow measurements are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.