Abstract
Critical parameters to be assessed in cell culture are the number of viable cells and cell viability. Growth, product formation, toxicity effects and the overall success of cell culture can depend largely on these. With interdigitated electrode structures (IDES) adherent cells are cultured directly on a pair of interdigitated electrodes, and the impedance of the system gives insight into the adhesive behaviour of the cells. The signal is influenced by the changes in number, growth and morphological behaviour of adherently growing cells, mainly owing to the insulating effects of the cell membranes. Five different cell lines are used, and their divergent behaviour is monitored over a period of four days, from inoculation of the cells to killing of the cells at the end of the experiments. Even when the cells from close monolayers, great fluctuations in the impedance signal can be observed. Nevertheless, for a more complete description of cellular systems, other parameters, such as acidification and respiration, have to be included in the measuring system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Medical & Biological Engineering & Computing
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.