Abstract
The development of a reliable, cost-effective instrument for on-line particle sizing remains a challenging area worldwide. This paper presents the design and evaluation of a new electrostatic sensor combined with a novel digital signal processing algorithm for the on-line continuous measurement of the mass median size of particles in a dilute-phase pneumatic suspension. Key design aspects of the constituent elements of the sensor, including a wire-mesh electrode, signal conditioning electronics and digital signal processing algorithm are described. Experimental results obtained using a typical range of particulate materials demonstrate that, under constant mass flow conditions, the processed output of the sensor is capable of indicating the mass median particle size with a relative error no greater than 15%. A critical appraisal of the technique, including advantages and limitations, is also included.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.