Abstract

We report the optimization, characterization, and validation of Adsorptive Square Wave Cathodic Stripping Voltammetry on antifouling gel-integrated microelectrode arrays for autonomous, direct monitoring of cobalt(II) metal species. Detection is accomplished by complexation with an added nioxime ligand. The limit of detection established for a 90 s accumulation time was 0.29 ± 0.01 nM in freshwater and 0.27 ± 0.06 nM in seawater. The microelectrode array was integrated in a submersible probe to automatically dose the complexing agent nioxime and realize an integrated sensing system. For the first time ever, the potentially bioavailable Co(II) fraction was determined in La Leyre River-Arcachon Bay continuum, enabling to evaluate the potential ecotoxicological impact of freshwater-carried Co(II) in the Arcachon Bay. The measured potentially bioavailable Co(II) concentrations were hazardous for aquatic biota along the continuum. The electrochemical Co(II) data were compared to ICP-MS data in various fractions to determine spatial Co(II) speciation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.