Abstract

The one-antenna method is adopted to the practical antenna validation needs in EMC laboratories. The proposed setup allows verification of important characteristics like antenna gain and antenna factor on a regular basis. The experiments show very good agreement with manufacturer calibration data. For frequency bands above 1 GHz the metallic wall of a typical shielded room can be used to calibrate e.g. wide band horn antennas. Considerations about potential sources of measurement errors are investigated, while the application of equi-ripple FIR filters - applied during the time domain gating - allows to minimize the uncertainty.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.