Abstract

In man, circulating leptin levels are increased with obesity and are regulated by a complex of hormonal, feeding and body-weight changes. Accurate and precise methods to quantitate circulating serum free leptin (f-leptin) concentrations are needed for physiological and clinical studies. We developed a one-step enzyme immunoassay to measure human f-leptin in serum. The detection limit was 0.40 ng/ml. The recovery of leptin added to serum was 90.8–102.8%. The within-run and between-day coefficients of variation (C.V.) ranged from 2.8 to 7.7 and 5.7 to 9.7%, respectively, and the immunoassay had an overall recovery rate for serial dilution in the range of 94.0–109.9%. Measured serum f-leptin concentrations in 201 adults correlated ( r=0.449, P<0.001) directly with body mass index (BMI kg/m 2), particularly when results were separated by gender ( r=0.709 for male, P<0.001; r=0.643 for female, P<0.001). We conclude that this one-step enzyme immunoassay is accurate for measuring f-leptin in human serum.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.