Abstract
Double-beam laser absorption detection in capillary electrophoresis (CE) has been developed. This is based on the direct subtraction of reference and signal photocurrents by an electronic circuit, under feedback control, to reduce background noise. A simple equation for calculating concentrations has been proposed and was confirmed by experimental results. A practical noise-to-signal ratio of 1 [times] 10[sup [minus]5] in intensity is achieved. This is 5 times lower than that of commercial CE systems. For absorbance detection, as low as 2 [times] 10[sup [minus]8] M malachite green can be detected. This corresponds to a 25-fold improvement of detection limit over commercial systems. This gain in detectability results from both a reduction in intensity fluctuations (noise) and an increase in the effective absorption path length (signal). 22 refs., 6 figs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.