Abstract

Magnetic bead-based lab-on-a-chip systems offer significant advantages compared to more conventional systems, mainly through the possibility of controlled manipulation of the magnetic carriers on-chip. In particular, microfluidic immunoassays using functionalized magnetic beads raise increasing interest. We present here a new approach for performing immuno-agglutination assays on-chip. Our system is based on a quadrupolar magnetic field set-up. Dynamic actuation of a confined plug of functionalized magnetic beads is used for analyte capture in a microchannel. A simple detection method based on the swelling of the released plug after agglutination is presented. We demonstrate the feasibility of on-chip agglutination tests by means of a streptavidin/biotinylated-bovine serum albumin (bBSA) model assay. A detection limit of about 200 pg/mL (approximately 3 pM) is achieved.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.