Abstract

Abstract The Knox equation, which relates the reduced plate height, h, to the reduced linear velocity, v, has a great importance in column and stationary phase testing. Significant A, B, and C terms of the Knox equation cannot be obtain with erroneous plate heights, h. Four methods for efficiency determination, the inflection (0.6H) method, the height/area method, an asymmetry based (0.1H) method and the moment method, were critically compared using 200 real chromatograms. Efficiency obtained using Gaussian assuming methods were highly overestimated, while the asymmetry based method gave quite acceptable results. Using the exponentially modified Gaussian model, it is demonstrated that the b/a ratio (the sum b + a is the peak-width at 10% of the peak height) is not an empirical figure of merit but can be related to a “peak skew” expression.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.