Abstract

Abstract: In this work, we evaluate the use of silicon photomultipliers (SiPMs) as a suitable candidate for use as a photodetector for thermoluminescence measurements. The measurements were based on the Hamamatsu S13360-series with a pixel area of 3 × 3 mm2 and microcell pitches of 25 μm, 50 μm, and 75 μm, which were mounted on a Riso TL/OSL reader system with LiF:Mg,Cu,Si chosen as the reference TL material. Through the experiment, we showed that the typical TL glow curves can be obtained with SiPMs, despite their lower geometric efficiencies. We also demonstrated the dose response of SiPMs ranging from 6 to 66 mGy for three different photon threshold levels. The detection limit was approximately equal to 0.2 mSv, 0.08 mSv, and 0.17 mSv for 25 μm, 50 μm, and 75 μm SiPMs, respectively. More importantly, a sensitivity change and sensor temperature depending on a TL reading up to 260 °C was monitored on a sequential readout. Lastly, we discuss some of the possible benefits and limitations of SiPM-based TL measurements in comparison to PMT-based TL measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.