Abstract
Correction for pile-up losses in the amplifier is possible by the dead-time fraction indicator of the ADC in case of long-lived radionuclides. If the dead-time meter has been calibrated, an accuracy of ∼1.5% is feasible up to a dead-time fraction of ∼25%. The precision decreases from ∼1.5% at ∼10% dead-time fraction to ∼3% at a deadtime fraction of 30%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Radioanalytical and Nuclear Chemistry Letters
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.