Abstract

In this paper, the thermal elevation in the human body due to the operation of a dual-unit epiretinal prosthesis to restore partial vision to the blind affected by irreversible retinal degeneration is presented. An accurate computational model of a 60-electrode device dissipating 97 mW power, currently under clinical trials is developed and positioned in a 0.25 mm resolution, heterogeneous model of the human head to resemble actual conditions of operation of the prosthesis. A novel simple finite difference scheme combining the explicit and the alternating-direction implicit (ADI) method has been developed and validated with existing methods. Simulation speed improvement up to 11 times was obtained for the the head model considered in this work with very good accuracy. Using this method, solutions of the bioheat equation were obtained for different placements of the implant. Comparison with in-vivo experimental measurements showed good agreement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.