Abstract

.Gamma-ray energy tracking arrays constitute the technological frontier of high-resolution gamma -ray spectroscopy revolutionizing modern nuclear physics investigations. Their principle of operation lies on the precise reconstruction of the three-dimensional gamma -ray interaction positions within the detector volume. The most common method to obtain these interaction points in real time is to compare the experimental signals against a reliable library of signals (signal basis) that maps the detector response as a function of the gamma -ray interaction position. Obtaining a high-fidelity signal basis, however, remains a big technological challenge, which hinders the optimal operation of these state-of-the-art arrays. In this article, we propose a pioneering and notably simple method for generating experimentally a reliable signal basis. The proposed method enables the gamma -ray tracking devices to perform a self-calibration of their position sensitive response in situ, opening up the way for reaching their optimum performance for the first time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.