Abstract

PurposeThe accuracy and precision of the dose estimates obtained with radiochromic film dosimetry are investigated in a clinical environment. The improvement in the accuracy of dose estimates reached with corrective methods is analyzed. Two novel re-calibration algorithms for radiochromic film dosimetry are presented. MethodsTwo different EBT3 lots are evaluated in two different centres. They are calibrated in Varian linacs and read in two different EPSON scaners. Once the lots are calibrated, three films per lot are considered and divided into stripes that are exposed to known doses. Several dosimetry protocols usually employed in radiochromic film dosimetry are used to convert film responses to absorbed doses. These protocols are characterized by different choices of the film responses or different sensitometric curves. Finally, the accuracy and reproducibility of the dose estimates is investigated with and without the corrective methods. Results and ConclusionsThe variabilities that affect radiochromic film dosimetry, such as intra-lot variability, inter-scan variability, post-exposure time and film autodevelopment may give rise to inaccuracies in the dose estimates. However, the implementation of re-calibration methods leads to more accurate dose estimates. All the investigated protocols showed more accurate and reproducible results when the re-calibrated methods were employed. So, the novel re-calibration methods may be applied in order to improve the accuracy and reproducibility of radiochromic film dosimetry.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call