Abstract
It is possible to obtain accurate results with the normal liquid scintillation method for the assay of radioactive paper strips, if the solubility of the radioactive compound being measured in the scintillation fluid is taken into account. A technique for taking this soiubility into account is described which involves the replacement of the radioactive paper strip in the sample container by an inactive paper strip and the use of a standard radioactive paper strip. The ratio of efficiencies for activity in solution and on the strip was determined for three different papers. (D.L.C.)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The International Journal Of Applied Radiation And Isotopes
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.