Abstract

In a pulse oximeter, clean artifact-free photoplethysmographic (PPG) signals with clearly separable DC and AC parts are necessary for error-free estimation of arterial oxygen saturation (SpO <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> ). Motion artifacts (MA) introduced in the PPG signals due to the movement of a patient result in a significant error in the readings of pulse oximeters and hence are a common cause of oximeter failure and loss of accuracy. In this paper, we present a comparative analysis of using different wavelets for the case of MA reduction from corrupted PPG signals. PPG signals with frequently encountered artifacts (horizontal, vertical and bending motions of finger) were recorded from the subjects and processed with different wavelets. Simulation results and statistical analysis revealed that the Daubechies wavelet performed well for obtaining clean motion artifact-free PPG signal. Interestingly, the wavelet based method proved to be efficient in reducing MA restoring the respiratory information in tact with the PPG.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.