Abstract

The mathematical model of flow injection analysis (FIA) for fast second-order chemical reaction in a straight tube is presented. The sample and reagent are initially premixed and the analytical solution for the detection output, i.e. the integral of the product of reaction concentration over the tube cross-section is given. The optimization of FIA is discussed. It is proposed to put the detector in the real FIA systems immediately after the point where the sample and reagent are satisfactorily mixed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.