Abstract
The MINOS light-injection calibration system has been fully documented in a previous article (Nucl. Instr. and Meth. A 492 (2002) 353). Upon commissioning in the MINOS detectors, the system was found to give a non-linear response to variations in the intensity of injected light. The source of this non-linearity has been traced to a small change in the spectrum of the injected light as a function of the current applied to the original blue LEDs, in combination with a rapidly varying spectral response function of the wavelength-shifting fibre used in the detector. Both aspects of the problem have been addressed successfully by use of LEDs with different spectral characteristics, and the system now has a linear response.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.