Abstract

The use of Euler’s homogeneous electric and magnetic fields can be a good basis for designing static mass spectrometers capable of operating in a spectrographic mode and enabling recording of mass spectra with a wide mass range in one measurement. It is proved mathematically that, when combined Euler’s homogeneous electric and magnetic fields are used to design a single-cascade mass-spectrometer, the double focusing mode cannot be achieved practically along the entire focal line. It is demonstrated that the use of mass spectrographic schemes in which the electric stage and the magnetic stage are separated is an inevitable solution in the creating of a mass spectrograph with high analytical performance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.