It is well known that photoacoustic tomography (PAT) can circumvent the photon scattering problem in optical imaging and achieve high-contrast and high-resolution imaging at centimeter depths. However, after two decades of development, the long-standing question of the imaging depth limit of PAT in biological tissues remains unclear. Here we propose a numerical framework for evaluating the imaging depth limit of PAT in the visible and the first near-infrared windows. The established framework simulates the physical process of PAT and consists of seven modules, including tissue modelling, photon transportation, photon to ultrasound conversion, sound field propagation, signal reception, image reconstruction, and imaging depth evaluation. The framework can simulate the imaging depth limits in general tissues, such as the human breast, the human abdomen-liver tissues, and the rodent whole body and provide accurate evaluation results. The study elucidates the fundamental imaging depth limit of PAT in biological tissues and can provide useful guidance for practical experiments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.