Abstract

AbstractThe best operating conditions that yield maximum sensitivity at minimum source pressure have been evaluated for the chemical ionization source in a Kratos MS‐25 mass spectrometer. The best reagent gas pressure is shown to be different for each gas and is significantly lower than is usually reported in the literature.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.