Abstract

X-ray imaging based on photon counting pixel detectors has received increased interest during the past years. Attached to a semiconductor of choice, some of these devices enable to resolve the spectral components of an image. This work presents the results from measuring the energy response function of a Medipix2 MXR Hexa detector, where six individual Medipix detectors were bump bonded to a 1 mm thick cadmium telluride sensor in order to form a 3×2 array of 4.2×2.8 cm 2 size. The average FWHM of the photo peak of an 241Am source was found to be 2.2 and 2.1 keV for single pixels and bias voltages of 200 and 350 V, respectively, across the whole Hexa detector. This corresponds to a relative energy resolution of less than 4%. Adding up all pixel spectra of individual chips lead to an only small deterioration of energy resolution, with line widths of 2.7 and 2.5 keV. In general, a lower detection efficiency was observed for the lower voltage setting, along with a shift of the peak position towards lower energies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.