Abstract

The linearity of dose response of chemical vapour deposition (CVD) diamonds grown at the Institute for Materials Research at Limburg University, Belgium, was investigated over a dose range relevant for radiotherapy. The following CVD diamonds were investigated: (1) a batch of square 3 x 3 mm2 detectors cut from a CVD wafer and (2) an as-grown CVD wafer of 6 cm diameter. A total of 20 CVD square detectors were irradiated with 137Cs gamma rays over the dose range from 200 mGy to 25 Gy. The CVD wafer, used as a large-area thermoluminescent (TL) detector, was exposed to a 226Ra needle. Very few square detectors showed linearity over a limited dose range, followed by saturation of the TL signal. The dose range of linearity was found to be strongly affected by the thermal annealing procedure of the detector. Owing to its high sensitivity and homogeneity of response, the large CVD diamond wafer was found to be very suitable as a large-area detector for 2-D dose mapping of the 226Ra brachytherapy source, possibly for Quality Assurance purposes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.