Abstract

In atom probe tomography (APT), multiple events can arise as a consequence of e.g. correlated field evaporation and molecular ion dissociation. They represent challenging cases for single-particle detectors and can cause compositional as well as spatial inaccuracies. Here, two state-of-the-art atom probe microscopes (Cameca LEAP 5000 XS and 5000 XR) were used to investigate cemented tungsten carbide, which exhibits high amounts of multiple events. By advanced data analysis methods, the natural character of the multiple events, as well as the performance of the APT detectors, are assessed. Accordingly, possible signal loss mechanisms are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.