Abstract

A linear algorithm for the rapid calculation of local rate constants is proposed. The method is applicable to the three-compartment models currently used in the analysis of positron camera measurements with [11C]methionine, [11C]deoxyglucose, and [11C]glucose. The same technique can also be used for the regional measurement of local blood flow with the aid of a freely diffusible tracer. The algorithm was applied to measurements on humans with [11C]glucose. As a comparison, the same data were also analyzed with a standard nonlinear technique. Good agreement between the two methods was obtained.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.