Abstract
Two image reconstruction methods currently dominate parallel MR imaging: SENSE and GRAPPA. While both seek to reconstruct images from subsampled multi-channel MRI data, there exist fundamental differences between the two. In particular, SENSE reconstructs an image of the excited spin-density directly whereas GRAPPA reconstructs estimates of the fully sampled raw coil data and then combines them to obtain an image. In this work we show that these differences can be exploited such that each method can compliment the other. In the case of SENSE, which requires an estimate of the coil sensitivity map before reconstruction, one can use GRAPPA to improve the coil sensitivity estimates. Alternatively, using coil sensitivity estimates and the SENSE reconstruction equations, one can improve the GRAPPA reconstruction parameter estimation. Together, these approaches can provide higher image quality than either method alone.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.