Abstract

The combination of short repetition times and large flip angles typically used in 3D contrast-enhanced magnetic resonance angiography (3D CE MRA) can significantly alter the expected shape of the slab profile for unenhanced tissues, which can cause increased aliasing in the slice select direction. In this work, this increased slice select aliasing is demonstrated and explained from both theoretical and experimental points of view. The effect is due to the Ernst angle of unenhanced background tissue occurring on the falling edges of the flip angle profile that has been set for the significantly reduced T1 of contrast-enhanced blood. The deleterious aliasing effects are magnified substantially when the chosen volume is placed close to surface coil reception with the slice select direction perpendicular to the coil axis. Magn Reson Med 44:336–338, 2000. © 2000 Wiley-Liss, Inc.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.