Abstract

Lattice-distortion analysis from high-resolution transmission electron micrographs offers a convenient and fast tool for direct measurement of strains in materials over a large area. In the present work, we have evaluated the accuracy of the strain measurement when the effects of the realistic experimental variables are explicitly taken into account by the use of image simulation techniques. These variables are focal setting and variation, local thickness and orientation of the sample, as well as misalignments of the sample and the incident beam. The evaluation reveals that consistency of image features and contrast within the micrographs is desired for the analysis to eliminate effects of the variations of local focus value and specimen thickness. After proper orientation of a crystalline specimen, the mis-orientation of the object will not notably influence the strain measurement even though a local bending may exist within the sample. However, the incident beam of the microscope needs to be aligned carefully as the beam misalignment may introduce a notable artefact around the interface region.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.