Abstract

In this study, a reliable and novel coupling of pipette-tip micro-solid phase extraction (PT-µSPE) with corona discharge ionization-ion mobility spectrometry (CD-IMS) was developed for on-site fast detection of benzodiazepines (BZDs) in dietary supplements. The poly (styrene-co-divinylbenzene) (St-co-DVB) monoliths fabricated in pipette-tips through an in-situ polycondensation reaction. The extraction procedure of PT-µSPE for aqueous sample was performed for an appropriate number of aspirating/dispensing cycles by using a manual micropipettor. Then analytes retained in the polymeric monoliths were eluted with elution solvents by repeating the aspirating/dispensing cycles, and the eluates were analyzed by CD-IMS. In order to find the best extraction conditions, several parameters such as the number of aspirating/dispensing cycles for extraction and desorption, pH, elution solvent, amount of polymer and ionic strength were investigated. Under the optimal conditions, the LOD of this method was in the range of 5–15 ng mL−1. The spiked recovery values in real sample were obtained by PT-µSPE-IMS in the range of 84.2–112.1% and validated by ultraperformance liquid chromatography tandem mass spectrometry (UPLC–MS/MS). PT-µSPE as an on-site sampling method can be suitably combined to IMS which has great potential to achieve high throughput detection by using a multichannel micropipette. The satisfactory analytical results reveal that the developed method can be effectively applied for on-site screening BZDs in health products.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call