Abstract

AbstractA poly(vinylchloride) (PVC) membrane based potentiometric immunosensor for the direct detection of alpha‐fetoprotein (AFP) has been developed. First, Au colloid particle was chemisorbed upon amino groups of o‐phenylenediamine, which were dissolved in plasticized PVC membrane. Then alpha‐fetoprotein antibody (anti‐AFP) was immobilized upon the surface of the Au colloid particle to prepare a potentiometric AFP immunosensor. The Au colloid particle modified PVC membrane was characterized by digital photo and transmission electron microscope (TEM). The immunosensor exhibited fast potentiometric response (≤4 min) and showed specific response to AFP in the range of 4.9 to 158.5 ng/mL with a correlation coefficient of 0.9971 and a detection limit of 1.6 ng/mL. The factors influencing the performance of the immunosensor were also studied in detail. Moreover, the proposed method is economical and efficient as well as potentially attractive for clinical immunoassays.