Abstract

ABSTRACT An on-line solid phase extraction system has been established for flow injection flame atomic absorption spectrometric determination of chromium on Chromotrope 2R coated Amberlite XAD-1180 (2R-1180) resin. Some parameters such as pH, flow rates of sample and elution solution, matrix effects on the quantitative recoveries of chromium from the resin were examined. The detection limit (3σ) was 2.59 µg/L for preconcentration at 6.5 mL/min. The method was applied for the determination of traces chromium in natural water samples with satisfactory results (relative standard deviations below 5%, recoveries >95%).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.