Abstract
A multi-reflection time-of-flight (MR-ToF) mass analyzer has been integrated into ISOLTRAP, the precision mass spectrometer for on-line mass determinations of short-lived nuclides at ISOLDE/CERN. The new instrument improves ISOLTRAP by providing a fast separation of isobaric contaminant species as well as subsequent ion selection using the fast Bradbury–Nielsen gate. Suppression ratios of up to 104 and mass-resolving powers of over 105 have been reached in off-line experiments. Preliminary data from on-line applications illustrate the benefit and performance of the device and its potential in the context of the ISOLTRAP setup.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.