Abstract
Heavy metal pollution from industrial wastewater is an important source. A method for heavy metals determination in industrial wastewater based on laser-induced breakdown spectroscopy (LIBS) technique was studied and the on-line monitoring system that used automatic graphite enrichment and spatial plasma confinement detection was developed and field demonstrated. The limits of detection (LOD) of heavy metal elements (Cd, Cr, Cu, Ni, Pb, Zn) could reach several μg/L. In Tongling, the on-line heavy metal monitor was field demonstrated. The calibration curves of copper and zinc were built on site, and then on-line monitoring was conducted. The measurement results of this monitor were compared with ICP-OES and had a good correlation. The results showed that the heavy metal monitor could be used for on-line detection of heavy metals in wastewater and had a good reliability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.