Abstract

A method using an automated on-line purge and trap gas chromatograph with a dry electrolytic conductivity detector (DELCD) has been developed for monitoring four regulated trihalomethanes in drinking water distribution systems. This analyzer samples trihalomethanes from drinking water by pervaporation through a silicone capillary membrane contained within a gas extraction cell (GEC) followed by preconcentration using an adsorbent trap. Trihalomethanes are subsequently desorbed from the trap onto a capillary column, separated and detected. The analyzer operates in real-time, samples directly from the drinking water distribution system and is fully automated. The optimization, operation, and evaluation of the analyzer and method are discussed. Method detection limits (MDL) are less than 1.0 μg L −1 with acceptable estimates for accuracy, and precision. The results from two on-line monitoring studies in chlorinated and chloraminated distribution systems are presented. The performance of the method is compared directly to United Stated Environmental Protection Agency Method 502.2 and shows a very slight, but acceptable bias.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.