Abstract
A new on-line preconcentration technique was developed that makes possible to determine nanomolar concentrations of weak acidic analytes in CE. The method consists of long-running electrokinetic sample injection and stacking (electrokinetic immobilization) of the analytes at a boundary of two electrolytes with different pH values (pH 9.5 and 2.5) and consequent mobilization of the stacked uncharged analytes in a micellar system (containing SDS micelles). Several factors including buffer concentration, pH, applied voltage, time of preconcentration, and SDS concentration were tested to optimize the analysis method. An about 4600-fold increase of the sample concentration (in comparison with the standard CZE) can be achieved during the preconcentration step. Two preservatives applied in food industry -- benzoic acid and sorbic acid were used as model samples. The applicability of the proposed method in food analysis was demonstrated by determination of nanomolar concentrations of benzoic acid in sunflower oil. An extended version of the computer program Simul was used for modeling both the preconcentration and mobilization processes taking place in the capillary.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.