Abstract

A MeV proton beam with a current in the range of fA–nA is usually measured using a secondary beam signal that is dependent on a sample or requires noise-free accurate charge integration in a Faraday cup. We propose a simple on-line beam current measurement set-up that can be used to directly measure beam current during ion beam analysis. A fast beam blanker is used to modulate the beam position after the last set of collimator slits in and out of a mini Faraday cup connected to fA-sensitive electrometer. Less than 100fA current can be measured during an undisturbed irradiation with kHz sampling frequency. The fast, simple and easily controlled current measurement set-up is going to be routinely used for ion beam analysis and modification at the new Lund sub-micron beam line.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.