Abstract

Surface plasmon resonance (SPR) is an optical technique that measures biomolecular interactions. Stand-alone SPR cannot distinguish different binding components present in one sample. Moreover, sample matrix components may show non-specific binding to the sensor surface, leading to detection interferences. This study describes the development of coupled size-exclusion chromatography (SEC) SPR sensing for the separation of sample components prior to their on-line bio-interaction analysis. A heterogeneous polyclonal human serum albumin antibody (anti-HSA) sample, which was characterized by proteomics analysis, was used as test sample. The proposed SEC-SPR coupling was optimized by studying system parameters, such as injection volume, flow rate and sample concentration, using immobilized HSA on the sensor chip. Automated switch valves were used for on-line regeneration of the SPR sensor chip in between injections and for potential chromatographic heart cutting experiments, allowing SPR detection of individual components. The performance of the SEC-SPR system was evaluated by the analysis of papain-digested anti-HSA sampled at different incubation time points. The new on-line SEC-SPR methodology allows specific label-free analysis of real-time interactions of eluting antibody sample constituents towards their antigenic target.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call