Abstract

Total elimination of the eluent resulting from the pre-separation step is the critical point when coupling LC to GC. As a helium flow is applied during transfer to eliminate the solvent, the interface used for linking the two chromatographic systems must be properly sealed to prevent gas leaks and to achieve an effective evaporation of the eluent. The aim of this work was to improve the performance of the Through Oven Transfer Adsorption Desorption (TOTAD) interface to remove the eluent coming from LC by modifying the way in which the injector sealing system is held in place. As with the original design, the new approach makes it possible to transfer high volumes at a high rate, but the proposed modification also simplifies the experimental work because the displacement risk of the sealing system is reduced. Analyses of an ester mixture by RPLC-GC were performed to confirm the applicability of the system modification. In this work, volumes of up to 5 ml, at flow rates as high as 2 ml/min, were transferred from LC to GC with almost complete solvent removal even when working in reversed phase mode in the LC step.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.