Abstract
A method for on-line, continuous measurement of the oxygen consumption rate ( Io 2) by mammalian cells on a microcarrier was developed and its reliability investigated. Utilizing the periodic dissolved oxygen (DO) fluctuation in the normal on-off control of DO, on-line, continuous measurement of Io 2 was carried out in which real-time estimation of the DO saturation concentration was made by measuring the gas-phase pressure and the gas-phase oxygen concentration. It was found that the continuously measured Io 2 value was quantitatively exact and could be applied commercially using the oxidation reaction of glucose by glucose oxidase.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.