Abstract

A simple and cost-effective laboratory-made liquid junction interface was used for coupling of CE with MS. In this device the capillary column and the spray tip were positioned in the electrode vessel containing appropriate spray liquid. The electrospray potential was applied on the electrode inside the liquid junction. A stable electrospray was produced at nanoliter per minute flow rates generated in the emitter tip without using an external pump. This arrangement provided high durability of the spray tip and independent optimization of the CE separation (use of coated capillaries) and ESI conditions. CE-MS analysis of mixtures of drugs, peptides, tryptic digests of proteins and biological fluids was optimized with respect to the effects of the distance between the separation capillary and electrospray tip and pressure applied on the liquid junction. The sensitivity of the system, in terms of the LOD (base peak monitoring) was below 10 ng/mL for the beta-blocker drugs and below 200 ng/mL for peptide analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.