Abstract

We use a terahertz echo pulse as a transfer standard to correct linear time-base errors of a terahertz spectrometer. With this calibration technique, the linear error of the translate stage is corrected and the spectral repeatability of the spectrometer is improved. This calibration technique is able to correct each ‘shot’ of the terahertz pulse and can realize on-line calibration of the terahertz spectrum. The average of multiple measurements of carbon monoxide gas absorption lines was corrected and the deviation was reduced by a factor of 2.5 using this technique. This on-line calibration technique needs no additional optical components and causes no performance degradation of the terahertz spectrometer. It provides a simple and alternative method to the conventional interferometric technique and allows for accurate measurement of spectral properties of specimens in the terahertz region.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.