Abstract

We have developed a densitometric method for measuring the isocentric accuracyand the accuracy of marking the isocentre position for linear acceleratorbased radiosurgery with circular collimators and room lasers. Isocentric shotsare used to determine the accuracy of marking the isocentre position with roomlasers and star shots are used to determine the wobble of the gantry and tablerotation movement, the effect of gantry sag, the stereotactic collimatoralignment, and the minimal distance between gantry and table rotation axes.Since the method is based on densitometric measurements, beam spot stabilityis implicitly tested. The method developed is also suitable for qualityassurance and has proved to be useful in optimizing isocentric accuracy. Themethod is simple to perform and only requires a film box and film scanner forinstrumentation. Thus, the method has the potential to become widely availableand may therefore be useful in standardizing the description of linearaccelerator based radiosurgical systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.