Abstract

We present for the first time a fully MEMS-integrated technology to manufacture 3D geometrically perfect solenoidal microcoils for microscale MRI applications. We report 25 microm isotropic resolution MR images of a copper sulfate aqueous phantom. These images are acquired using microcoils with 5 windings of insulated 25 microm diameter Au wire and with quality factors as high as 46 at the operating frequency (400 MHz).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.