Abstract

For the first time DNA coils formed by rolling circle amplification are quantified on-chip by Brownian relaxation measurements on magnetic nanobeads using a magnetoresistive sensor. No external magnetic fields are required besides the magnetic field arising from the current through the sensor, which makes the setup very compact. Limits of detection down to 500 Bacillus globigii spores and 2 pM of Vibrio cholerae are demonstrated, which are on the same order of magnitude or lower than those achieved previously using a commercial macro-scale AC susceptometer. The chip-based readout is an important step towards the realization of field tests based on rolling circle amplification molecular analyses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.