Abstract
Flow cytometry analysis requires a large amount of isolated, labelled, and purified cells for accurate results. To address the demand for a large quantity of cells prepared in a timely manner, we describe a novel microfluidic trap structure array for on-chip cell labelling, such as intracellular and extracellular labelling, and subsequent washing and release of cells. Each device contains [Formula: see text] trap structures, which made the preparation of large numbers of cells [Formula: see text] possible. The structure has a streamlined shape, which minimizes clogging of cells in capture and release steps. The trap structure arrays are built and tested using leukocytes, with different load flow speeds, incubation times, and release flow speeds. ∼85% of cells are captured independent of the input flow speed. The release efficiency depends on the incubation time, with over ∼80% of captured cells released for up to 20 min incubation, and on-chip labelling and washing with STYO13 are demonstrated. Qualitative models are developed as guidance for designing the proposed trap structure and to explain the increased performance over previous approaches.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.