Abstract
An expression relating the constant A to the characteristics of an ionisation chamber is developed. This constant, used to obtain the mass ionisation from the instrument reading, appears in expressions for CE and Clambda , and is a function of the chamber wall material. Measurements with a Baldwin-Farmer chamber are described, and indicate that the effective wall material is more nearly equivalent to Lucite than to air in 60Co gamma radiation. The resulting value of A is nearly 1.5% lower than the figure commonly used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.