Abstract
We present a method for calibrating piezoelectric sensors using a laser Doppler vibrometer. Our method uses an average of Fourier transform terms of the recorded signal from the piezoelectric sensor, which is compared with the laser probe measurement in the overlapping frequency range. We use our method to calibrate the response of miniature needle sensors employed in acoustic emission testing to several different excitation sources of stress waves in the frequency range of 20-300 kHz. We demonstrate that the output of the piezoelectric sensors can be accurately scaled with particle velocity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.