Abstract
Digital microfluidics is an emerging class of lab-on-a-chip system. Reliability is a critical performance parameter as these biochips are employed in various safety-critical biomedical applications. With the introduction of highly scalable, reconfigurable and field programmable Micro-Electrode-Dot-Array (MEDA) architecture, the limitation of conventional DMFBs in varying the droplet size/volume in fine grain manner has been resolved. However, the MEDA-based biochips must be adequately tested upon fabrication to guarantee the correctness of bioassays. In this work, an offline testing approach based on Oscillation-Based Testing (OBT) methodology is presented for MEDA-based digital microfluidic biochips. Various simulations were performed for droplet-electrode short fault model involving single and multiple micro-electrodes. Furthermore, the loss of droplet volume due to the presence of defect was analyzed using COMSOL Multiphysics. The simulation results based on PSpice and COMSOL show that the proposed approach is effective for detecting defects in MEDA-based biochips.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.